Draft - 1 Aug 08
Understanding SpO
2
Alarms
SpO
2
Monitoring
8-21
below the Low Limit. The SpO
2
high alarm limit can be set between 51 and
100% inclusive, in 1% increments.
The delay between the physiologic alarm condition and alarm annunciation at
the Information Center is <16 seconds. This means that the Information Center
will generate an alarm if the averaged numeric value on the display exists
beyond the alarm limit for more than a maximum of 16 seconds.
Setting the high SpO
2
alarm limit to 100% is equivalent to switching off the
high alarm. Therefore the upper alarm limit for oxygen saturation must be
carefully selected in accordance with accepted clinical practices.
The default setting for SpO
2
yellow alarms is latched. That is, when an SpO
2
limit is exceeded, you will need to acknowledge it at the Information Center.
The sound will be silenced but the message will remain on the display until the
condition is resolved.
See Chapter 5, “Alarms” for a list of the SpO
2
alarms.
Содержание IntelliVue TRx
Страница 6: ...Draft 1 Aug 08 About this Book vi ...
Страница 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Страница 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Страница 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Страница 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Страница 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Страница 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Страница 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Страница 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Страница 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Страница 274: ...Draft 1 Aug 08 Index 4 ...