ST/AR QT Interval Algorithm
ST/AR Arrhythmia Monitoring
7-17
QT Alarms
Alarm
Definition
**QTc High
The QTc high limit alarm is a long yellow alarm that
occurs when there are two consecutive 5 minute QTc
values above the set alarm limit.
**dQTc High
The dQTc alarm is a long yellow alarm that occurs
when the difference between the current value and
the baseline value exceeds the set limit for two
consecutive 5 minute dQTc values. The dQTc alarm
has a lower priority than the QTc High alarm.
Cannot Analyze QT
When there are two consecutive invalid 5 minute
values the Information Center generates a Cannot
Analyze QT soft INOP and displays a question mark
(?) for the QT parameter value. The Information
Center displays the Cannot Analyze QT INOP
without the question mark during initial startup and
during the learning phase.
Содержание IntelliVue TRx
Страница 6: ...Draft 1 Aug 08 About this Book vi ...
Страница 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Страница 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Страница 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Страница 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Страница 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Страница 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Страница 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Страница 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Страница 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Страница 274: ...Draft 1 Aug 08 Index 4 ...