Draft - 1 Aug 08
ST/AR ST Segment Algorithm
ST/AR Arrhythmia Monitoring
7-13
Adjusting ST
Alarms
Make adjustments to ST alarms on the ST Alarms window at the Information
Center.
Step
Action
1
From the Patient Window, select the
All Controls
button.
2
From the All Controls window, select the
ST Alarms
button under
Alarm Management and Setup.
3
In the ST Alarms window, adjust alarms as needed. Choices for
setting the ST alarm limits are:
Unit Settings—
Click on this button if want to have the specific
limits that are pre-set for your unit.
Smart Limits—
Click on this button to set high and low limits around
your patient's current ST value. The difference above and below the
patient's ST value are pre-set for your unit.
Note—
Smart Limits can be configured to automatically be activated
when the patient is connected. See the
IntelliVue Information Center
Instructions for Use
for additional information on using smart limits.
Set the high and low alarm limits based on your assessment of the
patient's clinical condition, unit protocols, or physician orders or
medication specified limits. A good guideline is + 1.0 mm or - 1.0
mm from the patient's ST, or follow your unit protocol.
Содержание IntelliVue TRx
Страница 6: ...Draft 1 Aug 08 About this Book vi ...
Страница 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Страница 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Страница 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Страница 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Страница 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Страница 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Страница 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Страница 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Страница 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Страница 274: ...Draft 1 Aug 08 Index 4 ...