Technical Alarms (INOPs)
Alarms
5-11
In the following table, technical alarms are listed alphabetically.
Technical Alarms (Inoperative Conditions)
Alarm Text
Priority
Condition
What to do
BATTERY LOW
BATTERY LOW T
Source - Telemetry
Soft
Battery power is low.
There is less than 15 minutes
of monitoring time
remaining. Replace batteries
promptly to avoid transceiver
shutdown and cessation of
monitoring.
CANNOT ANALYZE
ECG
Source - Information
Center
Hard
Arrhythmia algorithm
cannot reliably analyze the
ECG data on any
monitored leads.
Assess the lead selections,
initiate relearn, and validate
analyzed rhythm. Check
other INOPs for possible
source of problem.
CANNOT ANALYZE QT
Source - Information
Center
Soft
There have been two
consecutive invalid 5
minute values.
See “Limitations for QT
Monitoring” on page 7-23.
CANNOT ANALYZE ST
Source - Information
Center
Soft
ST algorithm cannot
reliably generate any valid
ST values on any
monitored lead.
Review the ECG signal
quality and correct if
necessary. Reposition the
ISO and J points.
Note—
If the patient has a
ventricular pacemaker, ST
analysis is not possible.
Содержание IntelliVue TRx
Страница 6: ...Draft 1 Aug 08 About this Book vi ...
Страница 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Страница 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Страница 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Страница 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Страница 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Страница 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Страница 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Страница 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Страница 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Страница 274: ...Draft 1 Aug 08 Index 4 ...