Draft - 1 Aug 08
Pairing Networked Devices
10-12
Pairing Monitoring Devices
Alarm
Behavior
(Networked
with Cable or
Short-Range
Radio
Connection)
The following table summarizes alarm behavior when a transceiver is paired
with an IntelliVue MP5, MP2, or X2 bedside monitor with a direct cable or
short-range radio connection.
Warning
Warning
As long as the ECG is being measured with the transceiver, there will be
no ECG signal available at the ECG analog output or ECG Sync Pulse
output.
Where Initiated
Effect at Bedside
Effect at Information Center
Silence
Bedside
Silences bedside and telemetry alarms.
Silences bedside and telemetry
alarms.
Information Center
Silences bedside and telemetry alarms.
Silences bedside and telemetry
alarms.