Draft - 1 Aug 08
Transceiver Operation when Connected to TeleMon
9-20
Telemetry Functions at the Information Center & TeleMon
SpO
2
Operation
•
SpO
2
is always in continuous mode.
•
Changes in SpO
2
mode do not take effect until after the transceiver is
disconnected from TeleMon. Mode settings are defined in the following
table:
•
If the sensor is disconnected from the transceiver during SpO2 monitoring
at TeleMon, a No Sensor technical alarm will be displayed at the
Information Center. Visual and auditory indicators for the No Sensor
Alarm are not active at Telemon.
Battery
•
The transceiver battery status is not available. When the Check button is
pressed, the battery gauge on the transceiver and at the Information Center
indicates full power, regardless of actual battery strength.
Defibrillation
•
In the event of patient defibrillation, it may take several seconds for the
ECG trace to reappear on the screen.
Telemetry Overview
•
If NBP is being monitored, NBP values will be displayed at the Overview
window.
Mode Selected at TeleMon
Mode at Disconnected Transceiver
Continuous
Continuous
5-min.
Continuous
1-min.
Continuous
Manual
Spot Check
Содержание IntelliVue TRx
Страница 6: ...Draft 1 Aug 08 About this Book vi ...
Страница 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Страница 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Страница 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Страница 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Страница 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Страница 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Страница 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Страница 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Страница 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Страница 274: ...Draft 1 Aug 08 Index 4 ...