Draft - 1 Aug 08
ST/AR ST Segment Algorithm
ST/AR Arrhythmia Monitoring
7-11
Establishing
ST Reference
Beats
(Baseline)
After adjusting the measurement points, you can establish baseline reference
beats for all available leads in the ST Review window at the IntelliVue
Information Center. Reference beats enable you to compare waveform changes,
for example from admission, or prior to or after treatment. The reference
continues to be saved beyond the 24 hour review window, but you can update it
to any beat within the last 24 hours. Please refer to the
IntelliVue Information
Center Instructions for Use or Online Help
for directions.
3
Adjust the J point, if necessary, by placing the cursor over the J-
point button to access the adjustment arrows. Then use the arrows to
position the bar at the end of the QRS complex and the beginning of
the ST segment.
4
Adjust the ST point, if necessary, by using the J point as an
“anchor” and placing the bar at the midpoint of the ST segment.
Choices are J+0, J+20, J+40, J+60, or J+80.
Step
Action
T
R
P
S
Q
J point
T
R
P
S
Q
ST point
Содержание IntelliVue TRx
Страница 6: ...Draft 1 Aug 08 About this Book vi ...
Страница 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Страница 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Страница 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Страница 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Страница 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Страница 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Страница 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Страница 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Страница 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Страница 274: ...Draft 1 Aug 08 Index 4 ...