Draft - 1 Aug 08
Positioning ECG Electrodes
ECG Monitoring
6-19
Lead
Placement
V1
on the fourth intercostal space at the right sternal
border
V2
on the fourth intercostal space at the left sternal
border
V3
midway between the V2 and V4 electrode
positions
V4
on the fifth intercostal space at the left
midclavicular line
V5
on the left anterior axillary line, horizontal with
the V4 electrode position
V6
on the left anterior axillary line, horizontal with
the V4 electrode position