Verifying Electrode Connections
6-26
ECG Monitoring
2
Expected Response
:
•
If 3-wire cable is attached: Red, White and Black indicators
illuminate, then all turn off.
•
If 5-wire cable in Standard mode is attached: Red, White,
Black, Green & Brown indicators illuminate, then all turn off.
•
If 5-wire cable in EASI mode is attached: Red, White, Black,
Green & Brown indicators illuminate, then all turn off. The
EASI indicator also illuminates briefly.
•
If 6-wire cable is attached: Red, White, Black, Green, Solid
Brown and Brown/White indicators illuminate, then all turn
off.
•
If all leads are off, the green indicator will be lit.
3
Unexpected Response
:
Any other response indicates a problem with the transceiver or lead
set.
•
Check the leadset connection. The colored line should not be
visible; otherwise the wrong leadset type may be detected.
•
If the INOP, INVALID LEADSET, and the corresponding
audible tone occur, the leadset inserted does not match the
transceiver’s configuration. Insert the correct leadset per the
configuration.
•
Use a new leadset.
If the problem is not corrected, contact your service provider.
Step
Action
Содержание IntelliVue TRx
Страница 6: ...Draft 1 Aug 08 About this Book vi ...
Страница 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Страница 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Страница 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Страница 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Страница 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Страница 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Страница 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Страница 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Страница 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Страница 274: ...Draft 1 Aug 08 Index 4 ...