Draft - 1 Aug 08
Cleaning and Sterilization
Maintenance, Cleaning & Troubleshooting
11-9
3
Apply -26 inHg +/- 1 (-12.77 psig +/- .49) vacuum to the empty
sterilizer chamber two times, to remove any residual EO or mois-
ture. Vent the vacuum pump to the outdoors to avoid toxic hazards
to personnel.
4
Insert the equipment to be processed into the gas sterilizer.
5
Heat the chamber and its contents to 54.4 +/- 2.8
o
C (130 +/- 5
o
F).
6
Apply -26 inHg +/- 1 (-12.77 psig +/- .49) vacuum to the sterilizer
chamber.
7
Humidify the chamber at 50% +/- 10% relative humidity for 20 to
30 minutes.
8
Taking a minimum of five minutes, slowly introduce EO sterilant
until the sterilizer unit pressure gauge reaches 11 +/- 1 psig.
Note—
At this pressure, the concentration of sterilant in the chamber
will be 600 +/- 50 mg/liter, regardless of the chamber size.
9
Process the equipment to be processed as follows:
Pressure: 11 +/- 1 psig (established in the preceding step).
Time:
2-3 hours
Temperature: 54.4 +/- 2.8
o
C (130 +/- 5
o
F)
Step
Action
Содержание IntelliVue TRx
Страница 6: ...Draft 1 Aug 08 About this Book vi ...
Страница 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Страница 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Страница 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Страница 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Страница 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Страница 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Страница 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Страница 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Страница 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Страница 274: ...Draft 1 Aug 08 Index 4 ...