Draft - 1 Aug 08
Measurement Specifications
12-16
Safety Standards & Specifications
Measurement Specifications
ECG
Parameter
Specification
ECG channel
transmitted Leads
•
3 electrodes
•
5 electrodes
•
5 electrodes,
EASI
•
6 electrodes
•
Channel #1 = I, II, or III
•
Channel #1 = II
Channel #2 = III
Channel #3 = MCL
•
Channel #1 = Va-i
Channel #2 = Va-s
Channel #3 = Ve-s
•
Channel #1= II
Channel #2 = III
Channel #3 = Va
Channel #4 = Vb
Resolution
5
μ
V
ECG Input
Differential, defibrillator protected against 360
joules discharge into a 100 ohm load
Input Impedance
> 5 megohms (@ 10 Hz
Input Dynamic Range
+/- 9 mV
DC Offset Range
+/- 320 mV
CMRR
> 90 dB @ 50, 60 Hz
Ban/- 3 dB
0.05 to 40 Hz
Содержание IntelliVue TRx
Страница 6: ...Draft 1 Aug 08 About this Book vi ...
Страница 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Страница 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Страница 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Страница 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Страница 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Страница 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Страница 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Страница 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Страница 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Страница 274: ...Draft 1 Aug 08 Index 4 ...