How the QT Analysis Algorithm Works
7-20
ST/AR Arrhythmia Monitoring
Select the QT Lead
The QT Lead field allows you to select which
leads to analyze when calculating the QT
parameters. To select the desired lead by
clicking on the QT Lead drop down arrow then
highlighting the lead from the list that displays.
Choose:
•
All
if you want a global QT measurement
based on all available leads. For standard
placement leads I, II, III, V and V1 through
V6 are used. For EASI placement directly
acquired AI, AS, and ES leads are used.
Note—
This may include leads that are not being
stored.
•
Primary
if you want to use the primary
lead for the QT measurement. If the primary
lead becomes unavailable or is changed the
QT measurement continues with the new
primary lead.
• A single lead from the list to use that lead
for QT measurement. If the lead you select
becomes unavailable QT monitoring stops.
Note—
The V7, V8, V9, V3R, V4R or V5R
leads are not available for single lead
selection. These leads are processed,
however, when you select
Primary
in the
QT Lead
field.
Adjustment
Description
Содержание IntelliVue TRx
Страница 6: ...Draft 1 Aug 08 About this Book vi ...
Страница 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Страница 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Страница 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Страница 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Страница 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Страница 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Страница 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Страница 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Страница 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Страница 274: ...Draft 1 Aug 08 Index 4 ...