Measurement Specifications
12-18
Safety Standards & Specifications
SpO
2
Parameter
Specification
SpO
2
Measurement
Range (Calibration and
Display)
0 to 100%
SpO
2
Accuracy
See table following.
SpO
2
Resolution
1%
SpO
2
Numerics -
Averaging
10 seconds
Note—
The update rate for the SpO
2
pulse
oximetry value and pulse rate is typically 1
second. This can be extended to a max. 60 s
when NIBP is measured on the same limb, with a
corresponding INOP message after a max. of 30
s, indicating that the displayed values are not
current values.
The effect of SpO
2
pulse oximetry on data
averaging is internally controllable by the
transmitter, with no user controls.
SpO
2
& Pulse Numerics
- Update Rate
Transmitted once per second.
Pleth Wave- Sampling
Rate
125 sps
Technical Alarms
(INOPs)
Triggered if the sensor is disconnected, if a pulse
is not detected, if the signal is noisy, if light
interference is detected, if the sensor is defective,
if the measurement is erratic, or if the module is
malfunctioning
Содержание IntelliVue TRx
Страница 6: ...Draft 1 Aug 08 About this Book vi ...
Страница 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Страница 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Страница 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Страница 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Страница 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Страница 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Страница 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Страница 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Страница 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Страница 274: ...Draft 1 Aug 08 Index 4 ...