Draft - 1 Aug 08
Measuring ECG
ECG Monitoring
6-9
Reconstructed
Leads
Reconstruction of leads from the sourced wave is defined by the calculations in
the following table. EASI reconstructed leads are a linear combination of all
three raw EASI leads.
ECG Lead
Clinical Calculations
in terms of electrodes
3-Wire
5-Wire
Standard
6-Wire
I
I
I
LA-RA
II
(default)
II
(default)
II
(default)
LL-RA
III
III
(default)
III
(default)
LL-LA
-
MCL
V-LA, where V=C
-
aVR
aVR
RA-(LA+LL)/2
-
aVL
aVL
LA-(RA+LL)/2
-
aVF
aVF
LL-(LA+RA)/2
-
V
(default)
V-(RA+LA+LL)/3, where V=C
Va
Va-(RA+LA+LL)/3, where
Va=V2 (default) position
Vb
Vb-(RA+LA+LL)/3, where Vb
=V5 (default) position
Содержание IntelliVue TRx
Страница 6: ...Draft 1 Aug 08 About this Book vi ...
Страница 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Страница 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Страница 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Страница 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Страница 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Страница 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Страница 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Страница 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Страница 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Страница 274: ...Draft 1 Aug 08 Index 4 ...