Draft - 1 Aug 08
Telemetry Functions at the Information Center
9-10
Telemetry Functions at the Information Center & TeleMon
Telemetry
Button
Determine the Information
Center response when Telemetry
Button is pressed.
Nurse Call
- generate nurse
call alarm
Record
- generate a recording
strip
Both
- generate nurse call
alarm and recording strip
None
Nurse Call
Telemetry
Device:
Volume
Set the volume level for all
adjustable sounds on the
transceiver.
1
(low) to
5
(high)
3
Telemetry
Device:
Mute
Enable/disable all adjustable
sounds on the transceiver.
enable
(unchecked) = sound
disable
(checked) = mute
mute
Standby
Duration
Sets the standby duration on the
device.
Infinite
10 minutes
20 minutes
30 minutes
1 hour
2 hours
3 hours
4 hours
Infinite
Enable
Remote
Suspend
Enable/disable alarm pause/
suspend at the transceiver.
enable
disable
disable
Unit Settings - Telemetry Setup
Control
Function
Settings
Factory
Default
Содержание IntelliVue TRx
Страница 6: ...Draft 1 Aug 08 About this Book vi ...
Страница 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Страница 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Страница 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Страница 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Страница 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Страница 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Страница 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Страница 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Страница 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Страница 274: ...Draft 1 Aug 08 Index 4 ...