Draft - 1 Aug 08
Measuring ECG
6-6
ECG Monitoring
There is no cardiotach within the transceiver; cardiotach analysis resides in the
arrhythmia algorithm at the Information Center. Therefore, arrhythmia analysis
is always turned on for telemetry patients. Arrhythmia analysis is either Basic or
Enhanced, depending on the product configuration.
ECG Leads
Monitored
Depending on the leadset connected to the transceiver, a different set of
viewable leads are available at the Information Center. The transceiver can
source up to four raw ECG waves.
If you are using ...
these leads can be selected at the
Information Center
3-wire
I, II, III
Sourced (raw) waves are received as:
•
Channel 1 = I, II, or III
Default is II.
5-wire (Standard mode)
I, II, III, aVR, aVL, aVF, MCL and V
Sourced (raw) waves are received as:
•
Channel 1 = II
•
Channel 2 = III
•
Channel 3 = MCL
Defaults are II, V, III.
Содержание IntelliVue TRx
Страница 6: ...Draft 1 Aug 08 About this Book vi ...
Страница 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Страница 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Страница 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Страница 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Страница 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Страница 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Страница 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Страница 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Страница 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Страница 274: ...Draft 1 Aug 08 Index 4 ...