Draft - 1 Aug 08
Cleaning and Sterilization
Maintenance, Cleaning & Troubleshooting
11-7
Perform the following steps to clean the transceiver of visible surface
contamination.
Note—
The cleaners listed above are also suitable for the optional Short Range
Radio Adapter (SRRA).
EO Sterilization
The transceiver can be subjected to EO sterilization four times per year for 2
years.
Equipment must first be cleaned (see “Cleaning the Transceiver” on page 11-5)
before this procedure is performed.
Note—
If there is concern over cross-contamination due to leadsets or sensors,
new leadsets or sensors should be used.
Caution
The Short Range Radio Adapter cannot be sterilized. Remove the adapter before
sterilizing the transceiver.
Step
Action
1
Remove the batteries and any cables or accessories.
2
Wipe the transceiver clean by using a cloth dampened modestly
with one of the approved cleaning agents listed in the table above.
3
If any visible residue remains in the Monitor/Service port, flush the
port with a forceful stream of water.
4
Rinse or wipe the transceiver with distilled water.
5
Allow to air-dry, or dry with a non-lint producing cloth.
Содержание IntelliVue TRx
Страница 6: ...Draft 1 Aug 08 About this Book vi ...
Страница 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Страница 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Страница 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Страница 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Страница 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Страница 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Страница 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Страница 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Страница 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Страница 274: ...Draft 1 Aug 08 Index 4 ...