Draft - 1 Aug 08
Physiologic Alarms
5-8
Alarms
* NON-SUSTAIN VT
Yellow
A run of Vs having a ventricular Heart
Rate greater than V-Tach limit but lasting
for less than the V-Tach Run limit.
ST/AR
Enhanced
Arrhythmia
* PACER NOT CAPT
Yellow
No QRS for 1.75 x the average R-R
interval with Pace Pulse (paced patient
only).
ST/AR
Basic &
Enhanced
Arrhythmia
* PACER NOT PACE
Yellow
No QRS and Pace Pulse for 1.75 x the
average R-R interval (paced patient only).
ST/AR
Basic &
Enhanced
Arrhythmia
* PAIR PVCs
Yellow
Two consecutive PVCs between non-
PVCs.
ST/AR
Enhanced
Arrhythmia
* PAUSE
Yellow
No QRS detected for x seconds. Choices
of >1.5 to 2.5 seconds.
ST/AR
Enhanced
Arrhythmia
* PVCs >xxx/MIN
Yellow
PVCs within one minute exceed by the
PVCs/min limit (xxx).
ST/AR
Basic &
Enhanced
Arrhythmia
** QTc High
Yellow
QTc value has exceeded the QTc High
limit for more than 10 minutes
ST/AR
Arrhythmia
**dQTc High
Yellow
dQTc High value has exceeded the dQTc
High Limit for more than 10 minutes.
ST/AR
Arrhythmia
Physiologic (Patient) Alarms
Alarm Text
Priority
Condition
Source
Содержание IntelliVue TRx
Страница 6: ...Draft 1 Aug 08 About this Book vi ...
Страница 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Страница 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Страница 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Страница 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Страница 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Страница 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Страница 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Страница 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Страница 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Страница 274: ...Draft 1 Aug 08 Index 4 ...