Draft - 1 Aug 08
Positioning ECG Electrodes
6-16
ECG Monitoring
5-Wire
Placement
(EASI Mode)
Lead
Corresponds
to Standard
Lead
Placement
E
V
on the lower sternum at the level of the
fifth intercostal space
A
LL
on the left midaxillary line at the same
level as the E electrode
S
LA
on the upper sternum
I
RA
on the right midaxillary line at the
same level as the E electrode
N
Reference
can be anywhere, usually below the
sixth rib on the right hip
Note—
Make sure that the S and E electrodes line up vertically on
the sternum, and that the I, E and A electrodes align horizontally.
E
S
A
I
Содержание IntelliVue TRx
Страница 6: ...Draft 1 Aug 08 About this Book vi ...
Страница 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Страница 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Страница 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Страница 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Страница 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Страница 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Страница 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Страница 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Страница 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Страница 274: ...Draft 1 Aug 08 Index 4 ...