Draft - 1 Aug 08
Cleaning and Sterilization
Maintenance, Cleaning & Troubleshooting
11-13
3
Test the transceiver:
a. Put fresh batteries in the transceiver (without a leadset
attached) and close the battery door
Result:
All six lead lights should flash, and one light
should remain on.
b. Attach a leadset to the ECG port, and attach an SpO
2
sensor to the SpO
2
port. If an ECG simulator is avail-
able, attach the ECG leads to the simulator and the SpO
2
sensor to yourself. At TeleMon, set the SpO
2
sample
rate to Continuous.
Result:
An ECG trace and SpO
2
information should be
visible on the Information Center display. All transceiv-
er lights should be off.
c. Disconnect the Right Arm lead for standard ECG or the
“I” electrode for EASI.
Result:
The RA LED or the “I” lead LED should turn
on, and a Leads Off INOP should appear on the display
at the Information Center.
d. Reconnect the electrode.
4
a. Connect the transceiver to TeleMon and observe the
ECG waveform and SpO
2
numerics on the TeleMon dis-
play.
Result:
The ECG waveform and SpO
2
numerics should
be displayed on the TeleMon screen.
Step
Action
Содержание IntelliVue TRx
Страница 6: ...Draft 1 Aug 08 About this Book vi ...
Страница 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Страница 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Страница 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Страница 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Страница 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Страница 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Страница 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Страница 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Страница 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Страница 274: ...Draft 1 Aug 08 Index 4 ...