Technical Alarms (INOPs)
Alarms
5-15
NO SIGNAL
Source - Information
Center
Hard
•
Transceiver is
outside the coverage
area, or
•
No batteries in
transceiver, or
•
Transceiver has
failed, or
•
Connection to
Database Server has
failed.
•
Short-range radio
connection has
failed.
•
Make sure that the
transceiver is within
the coverage area and
has good batteries.
•
Replace the transceiver
if Power On Self Test
fails.
•
Put bed in Standby.
•
Contact Service.
OUT OF AREA
Source - Information
Center
Hard
The transceiver is
communicating with an
access point designated as a
boundary limit.
For more information only.
REPLACE BATTERY
REPLACE BATTERY T
Note—
This INOP may also
be configured to display as
a Red or Yellow Technical
Alarm.
Source - Telemetry
Red or
Yellow or
Hard
Technical
Alarm,
Latched
Dead battery. No
monitoring is occurring.
Replace batteries.
SOME ECG ALRMS OFF
Source - Information
Center
Soft
Some yellow arrhythmia
alarms have been turned off
for this patient.
For information only.
Technical Alarms (Inoperative Conditions)
Alarm Text
Priority
Condition
What to do
Содержание IntelliVue TRx
Страница 6: ...Draft 1 Aug 08 About this Book vi ...
Страница 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Страница 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Страница 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Страница 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Страница 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Страница 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Страница 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Страница 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Страница 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Страница 274: ...Draft 1 Aug 08 Index 4 ...