M3500B/M5500B Service Manual
3
-
13
System Messages
3
System Messages
System messages remain on the display until the specified action is taken or
no longer relevant. They are intended for the end user and appear in the
Instructions for Use
. They are duplicated here for the reference of the service
person.
Before replacing any parts, check to see if the cables and flex circuits are
properly connected. See "Servicing Notes" on page 4-1.
Table 3-3 System Messages
Message
Description
Corrective Action
Attach Pads Cable
The pads cable is not properly attached to
the device.
Check the cable connection
Configuration Lost
The configuration is reset to the default
settings.
1. Reconfigure the M3500B/M5500B.
2. If problem reoccurs, service the unit.
See "Troubleshooting" on page 3-1 and
Table 3-6 on page 3-18.
Data Card Disabled
The PC card is not in use because it is full,
incompatible, absent, or inserted after the
M3500B/M5500B was turned on.
If possible, turn the M3500B/M5500B off
for more than 2 minutes, then insert a
new, empty, M3510A/M5510A Data Card
and turn the device on.
ECG Fault
The ECG data acquisition system failed
and data is unavailable from the 3- or 5-
lead monitoring electrodes.
Service the unit. See "Troubleshooting" on
page 3-1 and Table 3-7 on page 3-19.
Monitor Failure - Cycle
Power
An error has occurred in the monitoring
subsystem.
Turn power off, then on. If message reap-
pears, service the unit. See "Trouble-
shooting" on page 3-1.
Low Battery
The battery has sufficient capacity remain-
ing to provide only about ten minutes of
monitoring time and six shocks before the
M3500B/M5500B shuts off.
Replace the battery with a fully charged
M3516A/M5516A battery.
Leads Off
The monitoring electrodes are not applied
or are not making proper contact with the
patient.
Check that the monitoring electrodes are
properly applied.
The ECG cable is not connected.
Check that the ECG cable is properly con-
nected.
The internal cable from the ECG connec-
tor to the Interface PCA is disconnected
Check the internal cable.
Failure on the Interface PCA.
Replace the Interface PCA.
No Pads
The multifunction defibrillation electrode
pads are not properly connected to the
M3500B/M5500B.
Check the pads cable connection.
Содержание Heartstart XLT
Страница 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Страница 2: ......
Страница 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Страница 6: ......
Страница 16: ......
Страница 20: ......
Страница 52: ......
Страница 110: ...Opening the Sealed Case 4 24 Removal and Replacement Figure 4 13 Battery Connector Location 4 Pin Battery Connector ...
Страница 195: ...M3500B M5500B Service Manual 6 17 System Functional Descriptions 6 Figure 6 3 Power Distribution and Fusing ...
Страница 205: ...M3500B M5500B Service Manual 7 9 Waveforms 150J continued 7 Waveforms 150J continued 150J 75 ohms 150J 100 ohms ...
Страница 206: ...Waveforms 150J continued 7 10 Specifications Waveforms 150J continued 150J 125 ohms 2ms div 150J 125 ohms 5 ms div ...
Страница 208: ...Waveforms 200J continued 7 12 Specifications Waveforms 200J continued 200J 75ohms 200J 100ohms ...
Страница 216: ......
Страница 223: ......
Страница 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...