M3500B/M5500B Service Manual
3
-
19
Operational Problems
3
Operational Problems
These tables describe problems that may arise while using the M3500B/
M5500B.
ECG Monitoring
The following table covers problems that might arise while monitoring ECG.
Table 3-7 Operational Problems - ECG Monitoring (Pads or Leads)
Symptom
Possible Cause
Suggested Solution
Noisy trace - constant noise on the
baseline.
Incorrect configuration - power line
frequency (50 or 60 Hz) or Filter set-
tings.
Check/change configuration as
needed.
Nearby source of constant interfer-
ence.
Try moving cable/leads; try relocat-
ing unit.
Failure in ECG front end.
Pads: Replace Power PCA.
Leads: Replace Interface PCA.
Failure in signal processing circuits.
Replace Control PCA.
Noisy trace - intermittent bursts of
noise or random spikes.
Nearby source of time-varying inter-
ference, such as radio transmitter
(paging, walkie-talkies) or X-Ray
system.
Try moving cable/leads; try relocat-
ing unit.
Failure in ECG front end.
Pads: Replace Power PCA.
Leads: Replace Interface PCA.
Failure in signal processing circuits.
Replace Control PCA.
Noisy trace - low frequency, peri-
odic.
More than one instrument con-
nected to the patient and active.
Only have one active instrument on
the patient at a time.
Flat line - no waveform, no
Leads
Off
or
Pads Off
message.
Short in internal ECG wiring or front
end.
Pads:
1. Replace Power PCA.
2. Replace Patient Con
its internal cable.
Leads:
1. Replace Interface PCA.
2. Replace ECG con
internal ECG cable.
Failure in signal processing circuits.
Replace Control PCA.
Содержание Heartstart XLT
Страница 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Страница 2: ......
Страница 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Страница 6: ......
Страница 16: ......
Страница 20: ......
Страница 52: ......
Страница 110: ...Opening the Sealed Case 4 24 Removal and Replacement Figure 4 13 Battery Connector Location 4 Pin Battery Connector ...
Страница 195: ...M3500B M5500B Service Manual 6 17 System Functional Descriptions 6 Figure 6 3 Power Distribution and Fusing ...
Страница 205: ...M3500B M5500B Service Manual 7 9 Waveforms 150J continued 7 Waveforms 150J continued 150J 75 ohms 150J 100 ohms ...
Страница 206: ...Waveforms 150J continued 7 10 Specifications Waveforms 150J continued 150J 125 ohms 2ms div 150J 125 ohms 5 ms div ...
Страница 208: ...Waveforms 200J continued 7 12 Specifications Waveforms 200J continued 200J 75ohms 200J 100ohms ...
Страница 216: ......
Страница 223: ......
Страница 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...