M3500B/M5500B Service Manual
4
-
51
Top Case Assemblies
4
Figure 4-29 Disconnecting the Interface PCA
b.
Carefully disconnect the flex circuit from the SpO
2
PCA to the Inter-
face PCA at the Interface PCA. This connector has a sliding bar latch
- slide out toward edge of PCA to release, then gently pull out flex as
shown in Figure 4-29.
c.
Disconnect the large ribbon cable from the Interface PCA to the Con-
trol PCA at the Interface PCA. Remove the metal retaining clip on
the top of the connector and pull the connector straight up.
d.
Disconnect the 2 flex circuits from the Manual keypad to Interface
PCA at Interface PCA. Pull up on top edge of white connector to
unlatch ZIF, then gently slide out flex.
Cable to ECG
Connector
Flex Circuit
to SpO2
PCA
Flex Circuits
to Manual
Keypad
Latch
Flex Circuits
to Main Key-
pad
Ribbon
Cable to
Control PCA
Metal Bail
T10 Screw
Содержание Heartstart XLT
Страница 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Страница 2: ......
Страница 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Страница 6: ......
Страница 16: ......
Страница 20: ......
Страница 52: ......
Страница 110: ...Opening the Sealed Case 4 24 Removal and Replacement Figure 4 13 Battery Connector Location 4 Pin Battery Connector ...
Страница 195: ...M3500B M5500B Service Manual 6 17 System Functional Descriptions 6 Figure 6 3 Power Distribution and Fusing ...
Страница 205: ...M3500B M5500B Service Manual 7 9 Waveforms 150J continued 7 Waveforms 150J continued 150J 75 ohms 150J 100 ohms ...
Страница 206: ...Waveforms 150J continued 7 10 Specifications Waveforms 150J continued 150J 125 ohms 2ms div 150J 125 ohms 5 ms div ...
Страница 208: ...Waveforms 200J continued 7 12 Specifications Waveforms 200J continued 200J 75ohms 200J 100ohms ...
Страница 216: ......
Страница 223: ......
Страница 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...