M3500B/M5500B Service Manual
4
-
49
Top Case Assemblies
4
b.
Peel the backing material off the rubber overlay. Handle carefully
and touch the adhesive as little as possible.
c.
Starting with the top edge, align the overlay exactly and roll it down
slowly into place.
d.
Press firmly all over, especially the edges, to adhere the new overlay
into place.
8. Replace the Manual Door.
9. Be sure to specifically check the operation of each of the keys during Per-
formance Verification by performing a Controls Test (see "To check the
Controls (keys):" on page 2-19).