M3500B/M5500B Service Manual
3
-
15
System Messages
3
SpO2 Noisy Signal
Excessive patient movement, electrical
interference, or optical interference is
present.
1. Minimize patient movement or apply the
sensor to a site with less movement.
2. Secure the sensor cable loosely to the
patient.
3. Reduce sources of electrical or optical
interference.
SpO2 Sensor Fail
The SpO
2
cable is not connected to the
device; or the cable or sensor are broken.
1. Attach the cable to the M3500B/
M5500B
.
2. Replace cable and/or sensor.
System Failure - Cycle
Power
A serious error has occurred.
Turn power off, then on. If message reap-
pears, service unit. See "Troubleshoot-
ing" on page 3-1.
Service Unit
Appears during the Shift/System Check
1. Replace Data Card.
2. Replace lithium battery.
3. Replace Control PCA.
Table 3-3 System Messages (Continued)
Message
Description
Corrective Action
Содержание Heartstart XLT
Страница 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Страница 2: ......
Страница 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Страница 6: ......
Страница 16: ......
Страница 20: ......
Страница 52: ......
Страница 110: ...Opening the Sealed Case 4 24 Removal and Replacement Figure 4 13 Battery Connector Location 4 Pin Battery Connector ...
Страница 195: ...M3500B M5500B Service Manual 6 17 System Functional Descriptions 6 Figure 6 3 Power Distribution and Fusing ...
Страница 205: ...M3500B M5500B Service Manual 7 9 Waveforms 150J continued 7 Waveforms 150J continued 150J 75 ohms 150J 100 ohms ...
Страница 206: ...Waveforms 150J continued 7 10 Specifications Waveforms 150J continued 150J 125 ohms 2ms div 150J 125 ohms 5 ms div ...
Страница 208: ...Waveforms 200J continued 7 12 Specifications Waveforms 200J continued 200J 75ohms 200J 100ohms ...
Страница 216: ......
Страница 223: ......
Страница 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...