Momentary Messages
3
-
18
Troubleshooting
Extended Self Test Failures
Should the unit report a
FAIL
in the Extended Self Test, resolve it using the
solutions below.
Table 3-6 Extended Self Test Failures
Failure
Possible Cause
Suggested Solution
RAM
ROM
System
CODEC
IRDA
Timebase
Failure on Control PCA.
Replace Control PCA.
System FAIL 10
Lithium backup battery failure.
Replace lithium battery.
Data Card
Data Card full, or incompatible, or
defective.
Replace Data Card with new, empty
M3510A/M5510A.
Failure on Control PCA.
Replace Control PCA.
Pacer
Defib
Failure on Power PCA.
Replace Power PCA.
Failure on Control PCA.
Replace Control PCA.
FE
Failure on Control PCA.
Replace Control PCA.
Failure on Power PCA.
Replace Power PCA.
Failure on Interface PCA.
Replace Interface PCA.
SpO2
Failure on SpO
2
PCA.
Replace SpO
2
PCA.
Failure on Control PCA.
Replace Control PCA.
Содержание Heartstart XLT
Страница 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Страница 2: ......
Страница 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Страница 6: ......
Страница 16: ......
Страница 20: ......
Страница 52: ......
Страница 110: ...Opening the Sealed Case 4 24 Removal and Replacement Figure 4 13 Battery Connector Location 4 Pin Battery Connector ...
Страница 195: ...M3500B M5500B Service Manual 6 17 System Functional Descriptions 6 Figure 6 3 Power Distribution and Fusing ...
Страница 205: ...M3500B M5500B Service Manual 7 9 Waveforms 150J continued 7 Waveforms 150J continued 150J 75 ohms 150J 100 ohms ...
Страница 206: ...Waveforms 150J continued 7 10 Specifications Waveforms 150J continued 150J 125 ohms 2ms div 150J 125 ohms 5 ms div ...
Страница 208: ...Waveforms 200J continued 7 12 Specifications Waveforms 200J continued 200J 75ohms 200J 100ohms ...
Страница 216: ......
Страница 223: ......
Страница 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...