Electromagnetic Compatibility
7
-
18
Specifications
Electromagnetic Compatibility
When using the M3500B defibrillator/monitor (with or without the M3517A
AC Charger), electromagnetic compatibility with surrounding devices should
be assessed.
A medical device can either generate or receive electromagnetic interference.
Testing for electromagnetic compatibility EMC) of the M3500B with and
without the appropriate accessories has been performed according to the inter-
national standard for EMC for medical devices (IEC 60601-1-2). This IEC
standard has been adopted in Europe as the European Norm (EN 60601-1-2).
The EMC standards describe tests for both emitted and received interference.
Emission tests deal with interference generated by the device being tested.
According to the EMC standards, the M3517A AC Power Module does not
generate interference.
WARNING
Radio frequency (RF) interference from nearby transmitting devices may degrade
performance of the M3500B/M5500B defibrillator/monitor. Electromagnetic com-
patibility with surrounding devices should be assessed prior to using the defibrilla-
tor.
Reducing Electromagnetic Interference
The M3500B defibrillator/monitor and M3517A AC Charger are susceptible
to interference from other RF energy sources and continuous, repetitive,
power line bursts. Examples of other sources of RF interference are medical
devices, cellular products, information technology equipment and radio/tele-
vision transmission. Should interference be encountered, as demonstrated by
artifact on the ECG or dramatic variations in SpO
2
values, attempt to locate
the source. Assess:
z
Is the interference intermittent or constant?
z
Does the interference occur only in certain locations?
z
Does the interference occur only when in close proximity to certain
medical devices?
z
Does the SpO2 value change dramatically when the M3517A AC
Charger is unplugged?
Once the source is located, attempt to attenuate the EMC coupling path by
distancing the defibrillator from the source as much as possible. If assistance
is needed, call your local service representative.
Restrictions for Use
Artifact on the ECG caused by electromagnetic interference should be evalu-
ated by a physician or physician authorized personnel to determine if it will
negatively impact patient diagnosis or treatment.
Содержание Heartstart XLT
Страница 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Страница 2: ......
Страница 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Страница 6: ......
Страница 16: ......
Страница 20: ......
Страница 52: ......
Страница 110: ...Opening the Sealed Case 4 24 Removal and Replacement Figure 4 13 Battery Connector Location 4 Pin Battery Connector ...
Страница 195: ...M3500B M5500B Service Manual 6 17 System Functional Descriptions 6 Figure 6 3 Power Distribution and Fusing ...
Страница 205: ...M3500B M5500B Service Manual 7 9 Waveforms 150J continued 7 Waveforms 150J continued 150J 75 ohms 150J 100 ohms ...
Страница 206: ...Waveforms 150J continued 7 10 Specifications Waveforms 150J continued 150J 125 ohms 2ms div 150J 125 ohms 5 ms div ...
Страница 208: ...Waveforms 200J continued 7 12 Specifications Waveforms 200J continued 200J 75ohms 200J 100ohms ...
Страница 216: ......
Страница 223: ......
Страница 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...