M3500B/M5500B Service Manual
4
-
55
Top Case Assemblies
4
Speaker
These instructions describe how to remove and replace the speaker assembly.
Removal
1. Disconnect the speaker from the Control PCA. Refer to Figure 4-18 to
identify the speaker connector.
2. Remove the 2 screws and clamps from the speaker. See Figure 4-30.
3. Peel up the tape holding the speaker wires in position. Lift the speaker
and its wires out of the case.
Figure 4-30 Removing the Speaker
Replacement
1. Place the new speaker into position in the case.
2. Install both clamps and both screws.
3. Lay the wires in the same position in the case as were the original wires,
and secure with the original tape (or with other plastic electrical tape).
4. Connect the speaker to the Control PCA.
Display Window
Tape
Speaker
Wires
Speaker
Clamp
Содержание Heartstart XLT
Страница 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Страница 2: ......
Страница 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Страница 6: ......
Страница 16: ......
Страница 20: ......
Страница 52: ......
Страница 110: ...Opening the Sealed Case 4 24 Removal and Replacement Figure 4 13 Battery Connector Location 4 Pin Battery Connector ...
Страница 195: ...M3500B M5500B Service Manual 6 17 System Functional Descriptions 6 Figure 6 3 Power Distribution and Fusing ...
Страница 205: ...M3500B M5500B Service Manual 7 9 Waveforms 150J continued 7 Waveforms 150J continued 150J 75 ohms 150J 100 ohms ...
Страница 206: ...Waveforms 150J continued 7 10 Specifications Waveforms 150J continued 150J 125 ohms 2ms div 150J 125 ohms 5 ms div ...
Страница 208: ...Waveforms 200J continued 7 12 Specifications Waveforms 200J continued 200J 75ohms 200J 100ohms ...
Страница 216: ......
Страница 223: ......
Страница 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...