The Language Support Tool
4
-
68
Removal and Replacement
z
If this is an SpO
2
upgrade, the serial number should already be
present. In this case, verify it against the factory-applied label on
the bottom of the case.
z
If this is a Control PCA replacement, program in the serial num-
ber found on the factory-applied label on the bottom of the case
using the softkeys as instructed on the screen. Be sure to program
it in accurately, as the serial number is used for all repair history
tracking.
5. Check all the displayed information carefully before proceeding.
z
If the displayed information is correct, follow the screen prompts to
save the configuration.
z
If any of the information is incorrect, follow the prompts to NOT save
the configuration, then start over by powering the unit off, then back
on.
6. Turn off the power. Wait 2 seconds.
7. Remove the Data Card by pressing the black eject button and pulling the
Data Card from the compartment.
8. Turn the unit back on and enter Configuration Mode (see "Test and
Inspection Matrix" on page 2-3).
9. Print the configuration and check it against the printout from before the
servicing began. Reset the configuration (or load it from a Data Card) as
needed.
10. Perform Performance Verification Testing as described in "Performance
Содержание Heartstart XLT
Страница 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Страница 2: ......
Страница 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Страница 6: ......
Страница 16: ......
Страница 20: ......
Страница 52: ......
Страница 110: ...Opening the Sealed Case 4 24 Removal and Replacement Figure 4 13 Battery Connector Location 4 Pin Battery Connector ...
Страница 195: ...M3500B M5500B Service Manual 6 17 System Functional Descriptions 6 Figure 6 3 Power Distribution and Fusing ...
Страница 205: ...M3500B M5500B Service Manual 7 9 Waveforms 150J continued 7 Waveforms 150J continued 150J 75 ohms 150J 100 ohms ...
Страница 206: ...Waveforms 150J continued 7 10 Specifications Waveforms 150J continued 150J 125 ohms 2ms div 150J 125 ohms 5 ms div ...
Страница 208: ...Waveforms 200J continued 7 12 Specifications Waveforms 200J continued 200J 75ohms 200J 100ohms ...
Страница 216: ......
Страница 223: ......
Страница 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...