Diagnostic Tests
2
-
26
Performance Verification and Safety Tests
7. Using the
button, increase the rate to 180 bpm.
8. Using the
button, increase the output to 200 mA.
9. The Pacer should be delivering a current of 200mA +20mA. Record the
delivered current indicated by the Pacer tester as "bbb".
10. The display on the M3500B/M5500B should read the delivered current as
measured by the Pacer tester ("bb") +20 mA.
11. Turn off Pacing by pressing the
button.