M3500B/M5500B Service Manual
7
-
7
Specifications
7
General
Dimensions:
4.05" (H) x 11.25" (W) x 13.65" (D).
103 mm x 286 mm x 347 mm.
Weight:
Standard Configuration weighs < 10 lbs (4.5 kg) including battery,
full roll of paper, defibrillator patient cable, 1 set of pads, and without the car-
rying case; with the carrying case, it weights < 14 lbs (6.4 kg).
Environmental
Temperature:
0
o
to 50
o
C operating, -20
o
to 70
o
C storage.
z
Thermal paper may darken above 55
o
C.
z
Charging the battery at temperatures above 35
o
C may degrade battery
life.
z
Storing the battery for extended periods at temperatures above 40
o
C
will reduce battery capacity and degrade battery life.
z
Operating and storage specifications for electrodes may vary. Refer to
the manufacturer’s specifications for details.
Humidity:
Up to 95% Relative Humidity
z
Printer paper may jam if paper is wet.
z
Thermal printer may be damaged if wet paper is allowed to dry while
in contact with printer elements.
Altitude:
z
Operating: up to 15,000ft.
z
Shipping: up to 20,000 ft.
Shock (drops onto concrete):
Unit survives 39" (1 m) drops onto all surfaces
(faces, corners, and edges) enclosed in carrying case. Exposed handle surfaces
survive 24" drops.
Vibration:
Mil Std. 810E 514.4 Category 6 Helicopter, General Storage,
UH60.
Water Resistance:
IPX4, splash proof per IEC 60529, with Data Card door
closed.
Safety:
Meets IEC 60601-1 (EN 60601-1), UL 2601-1,
CAN/CSA C22.2 NO.601.1
EMC:
Meets EN 60601-1-2.
Other Considerations:
Equipment not suitable for use in the presence of a
flammable anesthetic mixture with air, oxygen, or nitrous oxide.
Mode of Operation:
Continuous.
Содержание Heartstart XLT
Страница 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Страница 2: ......
Страница 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Страница 6: ......
Страница 16: ......
Страница 20: ......
Страница 52: ......
Страница 110: ...Opening the Sealed Case 4 24 Removal and Replacement Figure 4 13 Battery Connector Location 4 Pin Battery Connector ...
Страница 195: ...M3500B M5500B Service Manual 6 17 System Functional Descriptions 6 Figure 6 3 Power Distribution and Fusing ...
Страница 205: ...M3500B M5500B Service Manual 7 9 Waveforms 150J continued 7 Waveforms 150J continued 150J 75 ohms 150J 100 ohms ...
Страница 206: ...Waveforms 150J continued 7 10 Specifications Waveforms 150J continued 150J 125 ohms 2ms div 150J 125 ohms 5 ms div ...
Страница 208: ...Waveforms 200J continued 7 12 Specifications Waveforms 200J continued 200J 75ohms 200J 100ohms ...
Страница 216: ......
Страница 223: ......
Страница 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...