M3500B/M5500B Service Manual
6
-
11
System Functional Descriptions
6
Pacing Functions
Pacing is initiated and controlled by pressing front panel keys. These key
presses are transferred from the keys to the Interface PCA via a flex circuit.
The key presses are detected and processed by the Interface PCA and then
passed to the Control PCA via a large ribbon cable.
The Control PCA directs the Power PCA to deliver the pacing pulses at the
rate and with the output selected by the user. The pacing pulses are delivered
via the pads cable to the defibrillation pads. The pacing current delivered is
reported back to the Control PCA, which sends the info to the display and
activates any printouts or screen messages as needed.
Содержание Heartstart XLT
Страница 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Страница 2: ......
Страница 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Страница 6: ......
Страница 16: ......
Страница 20: ......
Страница 52: ......
Страница 110: ...Opening the Sealed Case 4 24 Removal and Replacement Figure 4 13 Battery Connector Location 4 Pin Battery Connector ...
Страница 195: ...M3500B M5500B Service Manual 6 17 System Functional Descriptions 6 Figure 6 3 Power Distribution and Fusing ...
Страница 205: ...M3500B M5500B Service Manual 7 9 Waveforms 150J continued 7 Waveforms 150J continued 150J 75 ohms 150J 100 ohms ...
Страница 206: ...Waveforms 150J continued 7 10 Specifications Waveforms 150J continued 150J 125 ohms 2ms div 150J 125 ohms 5 ms div ...
Страница 208: ...Waveforms 200J continued 7 12 Specifications Waveforms 200J continued 200J 75ohms 200J 100ohms ...
Страница 216: ......
Страница 223: ......
Страница 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...