System Functional Descriptions
6
-
18
Theory of Operation
Lithium Backup Battery
The Backup Battery (located on the Control PCA) provides standby power to
maintain data during times when the main battery is either absent or dis-
charged and no external power is supplied. The data maintained includes the
user’s configuration choices, and the system time and date.
Data Card
The Data Card allows capture of key information such as ECG waveform,
shock advisories, charging, and delivering a shock. The Data Card is read by
the Event Review Data Management system for post-event analysis.
CAUTION
The Data Card must only be inserted or removed when the unit’s power is off.
A Data Card inserted while the power is on will not be recognized by the unit.
If a Card is removed while the power is on, its stored data may be corrupted
and rendered unreadable.
Содержание Heartstart XLT
Страница 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Страница 2: ......
Страница 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Страница 6: ......
Страница 16: ......
Страница 20: ......
Страница 52: ......
Страница 110: ...Opening the Sealed Case 4 24 Removal and Replacement Figure 4 13 Battery Connector Location 4 Pin Battery Connector ...
Страница 195: ...M3500B M5500B Service Manual 6 17 System Functional Descriptions 6 Figure 6 3 Power Distribution and Fusing ...
Страница 205: ...M3500B M5500B Service Manual 7 9 Waveforms 150J continued 7 Waveforms 150J continued 150J 75 ohms 150J 100 ohms ...
Страница 206: ...Waveforms 150J continued 7 10 Specifications Waveforms 150J continued 150J 125 ohms 2ms div 150J 125 ohms 5 ms div ...
Страница 208: ...Waveforms 200J continued 7 12 Specifications Waveforms 200J continued 200J 75ohms 200J 100ohms ...
Страница 216: ......
Страница 223: ......
Страница 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...