System Description
74
analysis time, the wavelengths required are sorted to minimize the change-over time
between any two consecutive wavelengths.
The dual monochromator system enables relatively high slits to be used with no loss
of image quality, which contributes to the high optical throughput. In addition, part
of the slit height is used for the simultaneous measurement of a neon reference
spectrum for wavelength correction.
Two slit width settings are available, optimized for UV and visible wavelengths. The
system automatically selects the correct slit for each analytical measurement.
Figure 3-2
Monochromator
Item
Description
Item
Description
1
Echelle Grating
4
Fiber optic for the neon reference
beam
2
Detector
5
Entrance Slit
3
Intermediate Slit
6
Prism
1
2
4
3
5
6
Содержание AVIO 200
Страница 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...Safety Practices 1 ...
Страница 32: ...30 Safety Practices ...
Страница 56: ...54 Safety Practices ...
Страница 57: ...Preparing Your Laboratory 2 ...
Страница 58: ...Preparing Your Laboratory 56 ...
Страница 70: ...Preparing Your Laboratory 68 ...
Страница 71: ...System Description 3 ...
Страница 72: ......
Страница 84: ...System Description 82 Figure 3 7 GemCone nebulizer and end cap N0680343 also shown in cross sec tional view ...
Страница 97: ...Installation 4 ...
Страница 98: ......
Страница 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Страница 164: ...Installation 162 ...
Страница 165: ...Maintenance 5 ...
Страница 166: ......
Страница 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Страница 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 272: ...Maintenance 270 ...
Страница 273: ...Troubleshooting 6 ...
Страница 274: ......
Страница 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Страница 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Страница 296: ...Troubleshooting 294 Figure 6 4 Aux Gas too high Air Leak or Spray Chamber Temperature too high ...
Страница 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Страница 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Страница 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Страница 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Страница 308: ......