Maintenance
226
9. If you have installed new 1/16-in. tubing, cut the tubing so that it extends
approximately 1/2 in. beyond the nut.
10. For the sample capillary tubing, use a piece of 0.023 in. (inner diameter)
polyethylene tubing (Part No. 09908265). Using a razor blade, bevel the tubing
by cutting it at an angle. Insert the sample capillary tubing into the tubing piece.
Removing the Argon Tip from the End Cap, N0780546
1. Loosen the nut on the argon inlet
.
Remove the nut along with the tubing and
ferrule. See the next figure.
2. Using your fingers, push the argon tip out from the inside of the end cap, being
careful not to damage it. Do
not
use any metal tools to remove the nebulizer tips
because this may damage the tips or end cap.
3. Inspect the argon tip. Clean the tip using soap and water or a dilute solution of
nitric acid. An ultrasonic bath may be used.
If you need to replace the tip, be sure to replace it with a
red
GemTip. GemTips
must be purchased in a set (Part No. N0780676). They are
not
available
separately.
4. Inspect the O-ring (Part No. N09921045), replace if it is worn or cracked.
Содержание AVIO 200
Страница 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...Safety Practices 1 ...
Страница 32: ...30 Safety Practices ...
Страница 56: ...54 Safety Practices ...
Страница 57: ...Preparing Your Laboratory 2 ...
Страница 58: ...Preparing Your Laboratory 56 ...
Страница 70: ...Preparing Your Laboratory 68 ...
Страница 71: ...System Description 3 ...
Страница 72: ......
Страница 84: ...System Description 82 Figure 3 7 GemCone nebulizer and end cap N0680343 also shown in cross sec tional view ...
Страница 97: ...Installation 4 ...
Страница 98: ......
Страница 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Страница 164: ...Installation 162 ...
Страница 165: ...Maintenance 5 ...
Страница 166: ......
Страница 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Страница 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 272: ...Maintenance 270 ...
Страница 273: ...Troubleshooting 6 ...
Страница 274: ......
Страница 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Страница 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Страница 296: ...Troubleshooting 294 Figure 6 4 Aux Gas too high Air Leak or Spray Chamber Temperature too high ...
Страница 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Страница 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Страница 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Страница 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Страница 308: ......