Maintenance
216
Nebulizers
PerkinElmer offers different types of nebulizers for a wide variety of applications.
For descriptions of the nebulizers and their different uses and advantages, refer to
Chapter 3, System Description. Part numbers and replacement parts for the different
types of nebulizers are listed on the next page.
Recommended flow rates for the nebulizers are given in the table below. If you find
that a flow rate is insufficient, this often indicates that the nebulizer has clogged and
requires cleaning. On GemTip and GemCone nebulizers, you should also check that
the argon gas fitting is properly tightened. If a space collar is used, tighten the fitting
with a wrench until you can no longer turn the knurled space collar. If a space collar
is not used, finger-tighten the fitting, then using a wrench, turn clockwise 1/8-turn.
Table 5-5.
Nebulizer Flow Rates
Argon Flow Rate
Nebulizer
Typical Value
Operating Range
GemTip Cross-Flow
0.6 L/min
0.4 - 1.3 L/min
Low-Flow GemCone
0.6 L/min
0.4 - 0.95 L/min
High Flow GemCone
1.0 L/min
0.7 - 1.0 L/min
Concentric Glass
(MEINHARD) Type A
0.70 L/min
0.4 - 1.3 L/min
Concentric Glass
(MEINHARD) Type C
0.70 L/min
0.4 - 1.3 L/min
Concentric Glass
(MEINHARD) Type K1
0.70 L/min
0.4 - 0.95 L/min
Concentric Glass
(MEINHARD) Type K3
0.70 L/min
2.5 - 3.5 mL/min
Ultrasonic
0.5 L/min
n/a
Mira Mist
0.70 L/min
0.6-1.0 L/min
Содержание AVIO 200
Страница 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...Safety Practices 1 ...
Страница 32: ...30 Safety Practices ...
Страница 56: ...54 Safety Practices ...
Страница 57: ...Preparing Your Laboratory 2 ...
Страница 58: ...Preparing Your Laboratory 56 ...
Страница 70: ...Preparing Your Laboratory 68 ...
Страница 71: ...System Description 3 ...
Страница 72: ......
Страница 84: ...System Description 82 Figure 3 7 GemCone nebulizer and end cap N0680343 also shown in cross sec tional view ...
Страница 97: ...Installation 4 ...
Страница 98: ......
Страница 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Страница 164: ...Installation 162 ...
Страница 165: ...Maintenance 5 ...
Страница 166: ......
Страница 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Страница 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 272: ...Maintenance 270 ...
Страница 273: ...Troubleshooting 6 ...
Страница 274: ......
Страница 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Страница 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Страница 296: ...Troubleshooting 294 Figure 6 4 Aux Gas too high Air Leak or Spray Chamber Temperature too high ...
Страница 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Страница 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Страница 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Страница 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Страница 308: ......