System Description
84
Peristaltic Pump
The peristaltic pump is fully computer-controlled. As an added feature, the pump
speed can be programmed to run at a fast speed for the read or rinse cycle. The
TubingSaver mode is a feature for extending the life span of pump tubing.
Autosampler
PerkinElmer offers autosamplers for automated sample handling. These
autosamplers consist of a sample table, a sample tray, and a motorized sampling arm
with an attached probe. Different sample trays are available for each autosampler,
covering a variety of sample volumes and total sample capacity requirements.
Concentric Glass
Nebulizer
(MEINHARD)
Provides excellent sensitivity and precision for aqueous solutions and
samples with few dissolved solids (less than 1%). Self-aspirating. Not to be
used with solutions containing hydrofluoric acid. PerkinElmer offers four
types, Meinhard (A, C, K1 and K3), which are described below. All require
an end cap (Part No. N0680343) for use with the Scott spray chambers.
The Meinhard nebulizer will aspirate liquid sample without pumping. The
nebulizer may also be operated with externally pumped sample, provided
that the sample flow is not significantly below the natural aspiration rate.
MEINHARD Type A
00472020
General purpose Meinhard nebulizer.
MEINHARD Type K1
N0777707
Low Ar Flow nebulizer for routine aqueous samples.
MEINHARD Type K3
N0681574
Low Ar Flow nebulizer for organic samples.
U-6000AT
+
Ultrasonic
Nebulizer
N0691709 (115 V)
N0691710 (230 V)
For samples with low analyte and low matrix concentrations. Typically
improves detection limits by a factor of 10 over conventional pneumatic
nebulizers.
Mira Mist Nebulizer
N0775330
The Mira Mist Nebulizer is recommended for aqueous solutions. Organics
will wet the nebulizer tip and performance will deteriorate.
Содержание AVIO 200
Страница 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...Safety Practices 1 ...
Страница 32: ...30 Safety Practices ...
Страница 56: ...54 Safety Practices ...
Страница 57: ...Preparing Your Laboratory 2 ...
Страница 58: ...Preparing Your Laboratory 56 ...
Страница 70: ...Preparing Your Laboratory 68 ...
Страница 71: ...System Description 3 ...
Страница 72: ......
Страница 84: ...System Description 82 Figure 3 7 GemCone nebulizer and end cap N0680343 also shown in cross sec tional view ...
Страница 97: ...Installation 4 ...
Страница 98: ......
Страница 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Страница 164: ...Installation 162 ...
Страница 165: ...Maintenance 5 ...
Страница 166: ......
Страница 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Страница 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 272: ...Maintenance 270 ...
Страница 273: ...Troubleshooting 6 ...
Страница 274: ......
Страница 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Страница 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Страница 296: ...Troubleshooting 294 Figure 6 4 Aux Gas too high Air Leak or Spray Chamber Temperature too high ...
Страница 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Страница 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Страница 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Страница 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Страница 308: ......