Maintenance
244
Cleaning the Nebulizer
For the longest life and best performance, wash your nebulizer by simply running
water as a sample for 10 minutes at the end of the day before shutting down the
plasma. Any other form of washing is usually unnecessary.
DO NOT
wash your
nebulizer in acid or solvents to "prevent salt build up".
DO NOT
use surfactants as
they will destroy the surface tension at the tip and performance.
Peristaltic Pump
The peristaltic pump requires routine maintenance to ensure that the system will run
properly. The pump tubing should be replaced when flat spots develop (on a daily
basis or after approximately eight hours of use). When replacing the pump tubing,
the drain tubing should be replaced at the same time since uneven wear can cause the
spray chamber to back up.
Replacement Pump Tubing
Replace the pump tubing when flat spots develop (after approximately eight hours
of use). When replacing the pump tubing, both the sample and drain tubing should
be replaced at the same time since uneven wear can cause the spray chamber to back
up.The following table gives part numbers for pump tubing for the sample and drain.
Caution
Attention
The gas orifice is at the very tip of the nebulizer. It is made of Teflon which
is
VERY SOFT
. This tip is very easily damaged and should
NEVER
be
touched with fingers, tissues, or anything else. If the tip is accidentally
touched and the nebulizer continues to operate, then it is still functional,
and its use can be safely continued.
L'orifice de gaz est à la pointe du nébuliseur. Il est fait de téflon qui est très
doux. Cette astuce est très facilement endommagé et ne doit
JAMAIS
être
touché avec les doigts, les tissus, ou toute autre chose. Si la pointe est
accidentellement touché et le nébuliseur continue à fonctionner, alors il est
encore fonctionnel, et son utilisation peut se poursuivre en toute sécurité.
Содержание AVIO 200
Страница 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...Safety Practices 1 ...
Страница 32: ...30 Safety Practices ...
Страница 56: ...54 Safety Practices ...
Страница 57: ...Preparing Your Laboratory 2 ...
Страница 58: ...Preparing Your Laboratory 56 ...
Страница 70: ...Preparing Your Laboratory 68 ...
Страница 71: ...System Description 3 ...
Страница 72: ......
Страница 84: ...System Description 82 Figure 3 7 GemCone nebulizer and end cap N0680343 also shown in cross sec tional view ...
Страница 97: ...Installation 4 ...
Страница 98: ......
Страница 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Страница 164: ...Installation 162 ...
Страница 165: ...Maintenance 5 ...
Страница 166: ......
Страница 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Страница 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 272: ...Maintenance 270 ...
Страница 273: ...Troubleshooting 6 ...
Страница 274: ......
Страница 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Страница 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Страница 296: ...Troubleshooting 294 Figure 6 4 Aux Gas too high Air Leak or Spray Chamber Temperature too high ...
Страница 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Страница 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Страница 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Страница 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Страница 308: ......