System Description
80
RF Control Electronics
The RF generator uses solid-state circuits. The solid state RF generator is designed
to significantly increase reliability and reduce the need for recalibration.
The RF generator also monitors plasma conditions. If the plasma is unstable, the
system automatically shuts it off.
To ensure operator safety and to protect the instrument from damage, the system
includes extensive use of RF shielding and safety interlocks. Proper RF shielding
and filtering are provided so that the system complies with regulations regarding
radio frequency radiation.
Sample Introduction System
Sample and Torch Compartments
The sample compartment has a front door to provide easy access to the torch
compartment with the sample introduction model inserting into the torch
compartment through the bottom of the compartment. The sample introduction
system is always visible and accessible even during operation. The spray chamber,
nebulizer, injector and torch can all be accessed underneath the torch compartment
without opening the door. The torch compartment door has a window so that the
operator can safely view the plasma and safety interlocks that shut down the plasma
if the door is opened during operation. The torch compartment has extensive
shielding to prevent exposure to radio frequency radiation.
Quick-Change Adjustable Torch Module
The Quick-Change Adjustable Torch Module consists of the quartz torch, injector,
spray chamber, and nebulizer/end cap – all in one assembly. This module can be
quickly removed from the sample compartment. This adjustable mount allows the
torch an adjustment of up to 5 mm.
Torch
The torch has a standard alumina injector with a 2.0 mm inner diameter and a
sapphire injector in 2.0 mm diameter. Other injectors available include alumina
injectors in different sizes for optimum sample loading of the plasma and quartz
injectors for different sample types.
Содержание AVIO 200
Страница 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...Safety Practices 1 ...
Страница 32: ...30 Safety Practices ...
Страница 56: ...54 Safety Practices ...
Страница 57: ...Preparing Your Laboratory 2 ...
Страница 58: ...Preparing Your Laboratory 56 ...
Страница 70: ...Preparing Your Laboratory 68 ...
Страница 71: ...System Description 3 ...
Страница 72: ......
Страница 84: ...System Description 82 Figure 3 7 GemCone nebulizer and end cap N0680343 also shown in cross sec tional view ...
Страница 97: ...Installation 4 ...
Страница 98: ......
Страница 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Страница 164: ...Installation 162 ...
Страница 165: ...Maintenance 5 ...
Страница 166: ......
Страница 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Страница 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 272: ...Maintenance 270 ...
Страница 273: ...Troubleshooting 6 ...
Страница 274: ......
Страница 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Страница 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Страница 296: ...Troubleshooting 294 Figure 6 4 Aux Gas too high Air Leak or Spray Chamber Temperature too high ...
Страница 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Страница 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Страница 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Страница 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Страница 308: ......