Maintenance
252
Adjusting the Pump Tubing
Adjust the tension on the sample and drain tubing, one at a time, using deionized
water as follows:
Adjusting the Sample Tubing
1. Make sure the sample tubing is centered in the channel, then place the clamp
around the tubing and lower the clamp arm on onto the pump. Turn the
adjustment screws until locked in place.
2. Place the sample capillary (SAMPLE IN side) in a container of deionized water.
3. Set the pump speed to 1.0 mL/min and start the pump.
4. Dip the tubing in DI water a few times to let air gaps form. As you watch the air
gaps form loosen the adjustment screw in a counterclockwise direction until the
air gaps stop or do not flow smoothly.
5. Tighten the adjustment screw
slowly
(turn clockwise) until droplets just start
forming at the end of the SAMPLE OUT side of the sample tubing.
6. Tighten the adjustment screw until you see a smooth flow of bubbles (3 turns on
the peristaltic pump). Add an additional three more turns.
7. Turn off the pump. The tension is now correctly set. No bubbles should be seen
in the sample tubing once the sample tube is kept in the solution.
Adjusting the Drain Tubing
1. Make sure the drain tubing is centered in the channel without disturbing the
sample tubing, then place the clamp around the drain tubing and lower the clamp
arm.
2. Loosen the adjustment screw (turn counterclockwise) for the drain tubing
channel so that the lever can be easily opened and closed with little resistance.
3. Set the pump speed to 1.0 mL/min and start the pump.
4. Dip the tubing in DI water a few times to let bubbles form. As you watch the
bubbles form loosen the adjustment screw in a counterclockwise direction until
the bubbles stop or do not flow smoothly.
Содержание AVIO 200
Страница 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...Safety Practices 1 ...
Страница 32: ...30 Safety Practices ...
Страница 56: ...54 Safety Practices ...
Страница 57: ...Preparing Your Laboratory 2 ...
Страница 58: ...Preparing Your Laboratory 56 ...
Страница 70: ...Preparing Your Laboratory 68 ...
Страница 71: ...System Description 3 ...
Страница 72: ......
Страница 84: ...System Description 82 Figure 3 7 GemCone nebulizer and end cap N0680343 also shown in cross sec tional view ...
Страница 97: ...Installation 4 ...
Страница 98: ......
Страница 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Страница 164: ...Installation 162 ...
Страница 165: ...Maintenance 5 ...
Страница 166: ......
Страница 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Страница 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 272: ...Maintenance 270 ...
Страница 273: ...Troubleshooting 6 ...
Страница 274: ......
Страница 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Страница 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Страница 296: ...Troubleshooting 294 Figure 6 4 Aux Gas too high Air Leak or Spray Chamber Temperature too high ...
Страница 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Страница 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Страница 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Страница 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Страница 308: ......