Installation
112
Connecting the System Components
The system component connections are as follows (see the following figure):
•
Ethernet cable from the computer to the instrument Ethernet connector on
the spectrometer side of the instrument. RS232 cable from an optional
autosampler to the computer.
•
The Avio 200 is equipped with an IEC 309 250 V 16/20A 2 pole plus
protective earth plug (PerkinElmer Part No. 09997530) that inserts into an
equivalent IEC 309 series receptacle (see Section 2, Electrical
Requirements).
•
Connect the computer and printer to electrical power as described in the
guides provided with these items.
Electrical Hazard
To prevent potential injury to yourself and damage to the instrument, switch
OFF all instruments in the system and disconnect them from the line power
supply before you perform the procedures described below.
Danger électrique
Pour éviter les risques de blessures pour vous et d'endommager l'appareil,
éteindre tous les instruments dans le système et les déconnecter de
l'alimentation de la ligne avant d'effectuer les procédures décrites ci-dessous.
Note
Extending Ethernet cables by making tandem connections will
compromise instrument communication.
Note
Make sure that the computer and printer share a common ground with the
instrument
.
!
Warning
Содержание AVIO 200
Страница 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...Safety Practices 1 ...
Страница 32: ...30 Safety Practices ...
Страница 56: ...54 Safety Practices ...
Страница 57: ...Preparing Your Laboratory 2 ...
Страница 58: ...Preparing Your Laboratory 56 ...
Страница 70: ...Preparing Your Laboratory 68 ...
Страница 71: ...System Description 3 ...
Страница 72: ......
Страница 84: ...System Description 82 Figure 3 7 GemCone nebulizer and end cap N0680343 also shown in cross sec tional view ...
Страница 97: ...Installation 4 ...
Страница 98: ......
Страница 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Страница 164: ...Installation 162 ...
Страница 165: ...Maintenance 5 ...
Страница 166: ......
Страница 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Страница 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 272: ...Maintenance 270 ...
Страница 273: ...Troubleshooting 6 ...
Страница 274: ......
Страница 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Страница 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Страница 296: ...Troubleshooting 294 Figure 6 4 Aux Gas too high Air Leak or Spray Chamber Temperature too high ...
Страница 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Страница 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Страница 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Страница 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Страница 308: ......