Maintenance
170
Nebulizer
Check the nebulizer spray pattern with deionized water. Clean or replace the
nebulizer as necessary.
Spray Chamber
Check for leaks around the nebulizer and drain fitting. Inspect the spray chamber for
deposits and check the condition of the O-rings.
Peristaltic Pump
Check that the pump rollers are clean, not scored, free from deposits due to spills,
and move freely. Remove and clean the pump head if necessary. Replace the pump
rollers if necessary.
Periodically observe the drain tubing. Liquid should drain smoothly from the spray
chamber. If not, adjust the pump tension on the pump tubing.
Drains
Check the spray chamber drain fittings, tubing and connections and replace if
necessary.
General System Maintenance
•
Air filters:
clean or replace the spectrometer and generator air filters as
necessary.
•
Chiller:
Flush out the chiller every 6 months. Fill the chiller reservoir with the
coolant fluid that includes inhibitor (Part No. N0776200). The chiller reservoir
has a capacity of 4.2 liters (1.1 US gallons).
Refer to Chapter 4, Installation for more information on filling the chiller.
Caution
Attention
Do not use deionized or tap water in the chiller. Only use coolant fluid
that includes inhibitor (Part No. N0776200).
Ne pas utiliser déminéralisée ou l'eau du robinet dans le refroidisseur.
Utiliser uniquement du fluide de refroidissement qui comprend un
inhibiteur (N0776200).
Содержание AVIO 200
Страница 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...Safety Practices 1 ...
Страница 32: ...30 Safety Practices ...
Страница 56: ...54 Safety Practices ...
Страница 57: ...Preparing Your Laboratory 2 ...
Страница 58: ...Preparing Your Laboratory 56 ...
Страница 70: ...Preparing Your Laboratory 68 ...
Страница 71: ...System Description 3 ...
Страница 72: ......
Страница 84: ...System Description 82 Figure 3 7 GemCone nebulizer and end cap N0680343 also shown in cross sec tional view ...
Страница 97: ...Installation 4 ...
Страница 98: ......
Страница 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Страница 164: ...Installation 162 ...
Страница 165: ...Maintenance 5 ...
Страница 166: ......
Страница 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Страница 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 272: ...Maintenance 270 ...
Страница 273: ...Troubleshooting 6 ...
Страница 274: ......
Страница 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Страница 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Страница 296: ...Troubleshooting 294 Figure 6 4 Aux Gas too high Air Leak or Spray Chamber Temperature too high ...
Страница 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Страница 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Страница 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Страница 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Страница 308: ......