Maintenance
228
5. To reassemble the nebulizer make sure that the O-ring is completely seated on
the base of the argon tip, then insert the argon tip into the argon inlet.
6. Check the condition of the argon inlet tubing (Part No. 09985723). If the tubing
is worn, replace it as follows. See the next figure.
a. Obtain a new piece of 1/8-in. O.D. Teflon tubing (Part No. 09985723).
b. Using a razor blade, make a straight cut on one end of the tubing. Wire cutters
or scissors do not make acceptable cuts.
c. Insert a nut (Part No. 09920545) and ferrule (Part No. 09920515) on the newly
cut end of the tubing, with the ferrule oriented as shown in the figure below.
Figure 5-25
Installing new tubing for the argon inlet. Cross-Flow End Cap, Part No.
N0780546.
7. With the tubing and ferrule assembled as shown in the figure above, screw the
nut into the argon inlet making a finger-tight connection. Do
not
use any tools
to tighten the nut.
8. Examine the two GemTips in the end cap. These tips should almost butt up to
each other at an angle of 90
.
Item
Description
1
Ferrule (Part No. 09920515)
2
Nut (Part No. 09920545)
3
1/8 inch O.D. Tubing (Part No. 09985723)
3
2
1
Содержание AVIO 200
Страница 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...Safety Practices 1 ...
Страница 32: ...30 Safety Practices ...
Страница 56: ...54 Safety Practices ...
Страница 57: ...Preparing Your Laboratory 2 ...
Страница 58: ...Preparing Your Laboratory 56 ...
Страница 70: ...Preparing Your Laboratory 68 ...
Страница 71: ...System Description 3 ...
Страница 72: ......
Страница 84: ...System Description 82 Figure 3 7 GemCone nebulizer and end cap N0680343 also shown in cross sec tional view ...
Страница 97: ...Installation 4 ...
Страница 98: ......
Страница 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Страница 164: ...Installation 162 ...
Страница 165: ...Maintenance 5 ...
Страница 166: ......
Страница 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Страница 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 272: ...Maintenance 270 ...
Страница 273: ...Troubleshooting 6 ...
Страница 274: ......
Страница 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Страница 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Страница 296: ...Troubleshooting 294 Figure 6 4 Aux Gas too high Air Leak or Spray Chamber Temperature too high ...
Страница 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Страница 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Страница 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Страница 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Страница 308: ......