Troubleshooting
276
•
Check that the injector is not clogged. You will need to remove the Quick-
Change Torch Module and disassemble the torch as described in
on page 173. Remove the spray chamber and either
remove the torch and then the injector or remove the injector leaving the torch
in place.
Performance Problems
Checking the Sample Introduction System
Many performance problems (such as poor precision or loss of signal) can be traced
to the setup and condition of the sample introduction system.
The following list describes general checks for the sample introduction system,
which should be done on a regular basis. Typical symptoms of problems are listed to
help you determine whether maintenance is required.
If you have already done performance checks, refer to the Syngistix Software Guide
on the Syngistix Document Pack CD for details
.
The sodium bullet test is also a good way of visually checking the sample flow in the
plasma and, therefore, indicates problems in sample introduction. This test is
described earlier in this chapter.
Pump and Pump Tubing
Is the tubing overstretched, or does it have flat spots, leaks, or discoloration?
Install new tubing, stretching it gently beforehand. If using organics, special tubing
should be used. See the list of replacement parts at the end of
Is the flow of liquid irregular?
Refer to the adjustment procedure in
Do the pump rollers bind or not roll easily? Are there grooves in the pump rollers?
On the PerkinElmer pump, the pump head may be removed for cleaning. The pump
head may also be replaced. See
Содержание AVIO 200
Страница 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...Safety Practices 1 ...
Страница 32: ...30 Safety Practices ...
Страница 56: ...54 Safety Practices ...
Страница 57: ...Preparing Your Laboratory 2 ...
Страница 58: ...Preparing Your Laboratory 56 ...
Страница 70: ...Preparing Your Laboratory 68 ...
Страница 71: ...System Description 3 ...
Страница 72: ......
Страница 84: ...System Description 82 Figure 3 7 GemCone nebulizer and end cap N0680343 also shown in cross sec tional view ...
Страница 97: ...Installation 4 ...
Страница 98: ......
Страница 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Страница 164: ...Installation 162 ...
Страница 165: ...Maintenance 5 ...
Страница 166: ......
Страница 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Страница 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 272: ...Maintenance 270 ...
Страница 273: ...Troubleshooting 6 ...
Страница 274: ......
Страница 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Страница 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Страница 296: ...Troubleshooting 294 Figure 6 4 Aux Gas too high Air Leak or Spray Chamber Temperature too high ...
Страница 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Страница 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Страница 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Страница 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Страница 308: ......