Maintenance
189
6. If needed, clean the torch, see
on page 184. If the torch is
cracked or broken carefully dispose of the torch.
7. To install the torch see
Installing the Quick-Change Adjustable Torch Module
Scott Spray Chamber
Check the spray chamber periodically for leaks around the nebulizer end cap and
drain fitting. Examine the end cap O-ring for cracking and wear and replace if
necessary. Deposits and plated out metals can accumulate in the spray chamber,
which can introduce “memory effects.” Usually, aspirating a strong acid or solvent
will clean out these deposits. If not, remove the spray chamber and clean it in a 5%
to 20% nitric acid bath
Removing the Scott Spray Chamber
1. Place the sample tubing in deionized water and turn on the pump to flush the
spray chamber. Remove the sample tubing from the deionized water and
continue to run the pump a few minutes until the spray chamber is fully drained.
Then turn off the pump.
2. Disconnect the spray chamber drain tubing from the pump tubing for the drain.
Be careful not to have liquid spill out of the disconnected drain tubing. Spilled
liquid should be cleaned up immediately. Check that the tubing is in good
condition and replace if necessary
When installing the Torch Module in the next steps, do so carefully so you will
avoid breaking the torch and risking possible injury.
Avertissement: Lors de l'installation du module de la flamme dans les
prochaines étapes, faire avec soin afin que vous éviter de casser la torche et un
risque de blessure possible.
Caution
Attention
Wear
cotton gloves
when you handle the torch glassware
.
Porter des gants de coton lorsque vous manipulez la verrerie de la torche.
!
Warning
Содержание AVIO 200
Страница 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...Safety Practices 1 ...
Страница 32: ...30 Safety Practices ...
Страница 56: ...54 Safety Practices ...
Страница 57: ...Preparing Your Laboratory 2 ...
Страница 58: ...Preparing Your Laboratory 56 ...
Страница 70: ...Preparing Your Laboratory 68 ...
Страница 71: ...System Description 3 ...
Страница 72: ......
Страница 84: ...System Description 82 Figure 3 7 GemCone nebulizer and end cap N0680343 also shown in cross sec tional view ...
Страница 97: ...Installation 4 ...
Страница 98: ......
Страница 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Страница 164: ...Installation 162 ...
Страница 165: ...Maintenance 5 ...
Страница 166: ......
Страница 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Страница 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 272: ...Maintenance 270 ...
Страница 273: ...Troubleshooting 6 ...
Страница 274: ......
Страница 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Страница 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Страница 296: ...Troubleshooting 294 Figure 6 4 Aux Gas too high Air Leak or Spray Chamber Temperature too high ...
Страница 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Страница 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Страница 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Страница 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Страница 308: ......