Maintenance
258
Changing the Neon Lamp Assembly
Tools Required:
T20 Torx driver
The neon lamp (Part No. B0810144) is in an assembly that plugs into a connector on
the right-hand side of the optics tub. A panel in the side of the spectrometer housing
gives access to the neon lamp. Refer to Figure 5-37.
1. Shut down and switch off the spectrometer.
2. Use the T20 Torx driver to remove the screw that secures the access panel on the
right-hand side of the spectrometer and remove the panel. Place the screw in a
secure location.
3. The neon lamp assembly is the metal box that is visible inside the access hole.
Loosen the large screw on the metal box inside the access panel and pull the
metal box off the side of the optics tub.
4. Carefully slide a new neon lamp assembly (Part No. B0810144) onto the tub
using the metal plates on the tub as guides. Tighten the large screw securely.
5. Replace the access panel and secure it with the screw.
Chiller and Filters
Inspect the metal strainer (Part No. N0776075) that is suspended inside the chiller
reservoir; this should be removed and rinsed clean when the system is flushed or if
dirty. Flush out the chiller every 6 months.
Fill the chiller reservoir with the coolant fluid with corrosion inhibitor (Part No.
N0776200). The chiller reservoir has a capacity of 4.2 liters (1.1 US gallons). Refer
to
Connecting the PolyScience Chiller
on page 104 for more information on filling
the chiller.
Always turn off the instrument power before this box is
plugged in or removed.
Avertissement:
Couper toujours l’alimnetation de
l’instrument avant que sette caisse soit branch
é
e ou retir
é
e.
!
Warning
Содержание AVIO 200
Страница 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...Safety Practices 1 ...
Страница 32: ...30 Safety Practices ...
Страница 56: ...54 Safety Practices ...
Страница 57: ...Preparing Your Laboratory 2 ...
Страница 58: ...Preparing Your Laboratory 56 ...
Страница 70: ...Preparing Your Laboratory 68 ...
Страница 71: ...System Description 3 ...
Страница 72: ......
Страница 84: ...System Description 82 Figure 3 7 GemCone nebulizer and end cap N0680343 also shown in cross sec tional view ...
Страница 97: ...Installation 4 ...
Страница 98: ......
Страница 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Страница 164: ...Installation 162 ...
Страница 165: ...Maintenance 5 ...
Страница 166: ......
Страница 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Страница 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 272: ...Maintenance 270 ...
Страница 273: ...Troubleshooting 6 ...
Страница 274: ......
Страница 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Страница 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Страница 296: ...Troubleshooting 294 Figure 6 4 Aux Gas too high Air Leak or Spray Chamber Temperature too high ...
Страница 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Страница 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Страница 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Страница 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Страница 308: ......