33
Instrument Safety Practices
kg (330 lbs) with the shipping container. If the instrument needs to be moved please
contact PerkinElmer Service.
Laboratory Ventilation
Toxic combustion products, metal vapor, and ozone can be generated by the
system, depending upon the type of analyses. Combustion products vary with the
type of sample analyzed. If the sample compartment door is opened, a maximum
of 0.9 liters of Argon may escape during the first minute. After that, 0.1 liters/
minute of Argon will continue to escape while the door is open.
An exhaust venting system is always required to remove those gases which are
generated during the normal operation of the instrument.
An efficient ventilation system must be provided for the instrument. Detailed
specifications for a recommended venting system are described in the section titled
Exhaust Vent Requirements on page 58.
Instrument Safety Practices
This section describes potential hazards and recommended safety practices. You
should thoroughly review this information.
This equipment requires no specified inspection or preventive maintenance
to ensure the continuous functioning of its safety features.
Do
not
position the instrument so that it is difficult to operate the main power
switch which is the disconnecting device.
Caution
Attention
The protection provided by this equipment may be impaired if
the equipment is used in a manner not specified by PerkinElmer.
La protection fournie par cet équipement risque d'être moins efficace
si l'équipement fait l'objet d'une utilisation différente de celle
mentionnée par PerkinElmer.
Содержание AVIO 200
Страница 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...Safety Practices 1 ...
Страница 32: ...30 Safety Practices ...
Страница 56: ...54 Safety Practices ...
Страница 57: ...Preparing Your Laboratory 2 ...
Страница 58: ...Preparing Your Laboratory 56 ...
Страница 70: ...Preparing Your Laboratory 68 ...
Страница 71: ...System Description 3 ...
Страница 72: ......
Страница 84: ...System Description 82 Figure 3 7 GemCone nebulizer and end cap N0680343 also shown in cross sec tional view ...
Страница 97: ...Installation 4 ...
Страница 98: ......
Страница 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Страница 164: ...Installation 162 ...
Страница 165: ...Maintenance 5 ...
Страница 166: ......
Страница 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Страница 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 272: ...Maintenance 270 ...
Страница 273: ...Troubleshooting 6 ...
Страница 274: ......
Страница 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Страница 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Страница 296: ...Troubleshooting 294 Figure 6 4 Aux Gas too high Air Leak or Spray Chamber Temperature too high ...
Страница 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Страница 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Страница 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Страница 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Страница 308: ......