Maintenance
172
GemTip or GemCone
Nebulizer/End Cap
Assembly
Clean with soap and water; rinse thoroughly.
Concentric Glass
(MEINHARD)
Nebulizer
Do
not
use an ultrasonic cleaner. Do
not
use a cleaning
wire. See
Concentric Glass Nebulizer
later in this
chapter.
Mira Mist Nebulizer
For the longest life and best performance, wash your
nebulizer by simply running water as a sample for 10
minutes at the end of the day before shutting down the
plasma. Any other form of washing is usually
unnecessary. Please
DO NOT
wash your nebulizer in
acid or solvents to 'prevent salt build up'.
DO NOT
use
surfactants as they will destroy the surface tension at
the tip and performance.
The tip is very easily
damaged and should NEVER be touched with
fingers, tissues, or anything else
.
Spray Chamber
By handling the spray chamber oil and dirt can
accumulate on it and this will affect performance. To
assure proper wetting wipe the spray chamber with
acetone to remove all oils and dirt.
Pump Head
Remove and clean with water or a mild solvent.
Purge Viewing
Windows/Tubes
Clean with deionized water and dry with a soft cloth.
To clean difficult stains, refer to the information in this
chapter.
Table 5-1.
Recommended Cleaning Procedure
(Continued)
Содержание AVIO 200
Страница 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...Safety Practices 1 ...
Страница 32: ...30 Safety Practices ...
Страница 56: ...54 Safety Practices ...
Страница 57: ...Preparing Your Laboratory 2 ...
Страница 58: ...Preparing Your Laboratory 56 ...
Страница 70: ...Preparing Your Laboratory 68 ...
Страница 71: ...System Description 3 ...
Страница 72: ......
Страница 84: ...System Description 82 Figure 3 7 GemCone nebulizer and end cap N0680343 also shown in cross sec tional view ...
Страница 97: ...Installation 4 ...
Страница 98: ......
Страница 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Страница 164: ...Installation 162 ...
Страница 165: ...Maintenance 5 ...
Страница 166: ......
Страница 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Страница 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 272: ...Maintenance 270 ...
Страница 273: ...Troubleshooting 6 ...
Страница 274: ......
Страница 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Страница 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Страница 296: ...Troubleshooting 294 Figure 6 4 Aux Gas too high Air Leak or Spray Chamber Temperature too high ...
Страница 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Страница 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Страница 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Страница 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Страница 308: ......