Maintenance
235
Installing the GemCone Nebulizer on the Cyclonic Spray Chamber
1. Before inserting the GemCone nebulizer into the chamber, connect the argon gas
fitting to the nebulizer threads. First, tighten the gas fitting by hand. Tighten the
fitting another 1/8 of a turn using a 9/16-in. open-end wrench. Support the
nebulizer with a back-up wrench when tightening or loosening fittings. Do
not
overtighten the fitting as this may cause leaking and poor ICP performance.
2. Connect the capillary tubing to the nebulizer. Be sure not to crimp or bend the
tubing as you insert it. Crimps may reduce flow to the tip of the nebulizer or may
cause air to enter the liquid stream during operation.
3. Insert the nebulizer into the cyclonic spray chamber as shown in the next figure.
The nebulizer may be inserted to its fullest extent.
Figure 5-28
Connecting the GemCone nebulizer to the cyclonic spray chamber.
Item
Description
Item
Description
1
Cyclonic Spray Chamber
4
GemCone Nebulizer Assembly
2
Tubing Assembly for the
GemCone Nebulizer (Part No.
N0770336)
5
Teflon Tubing-1 inch (Part No.
09985708)
3
Argon Gas Fitting
6
Capillary Tubing (Part No. 09908265)
1
2
3
5
6
4
Содержание AVIO 200
Страница 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...Safety Practices 1 ...
Страница 32: ...30 Safety Practices ...
Страница 56: ...54 Safety Practices ...
Страница 57: ...Preparing Your Laboratory 2 ...
Страница 58: ...Preparing Your Laboratory 56 ...
Страница 70: ...Preparing Your Laboratory 68 ...
Страница 71: ...System Description 3 ...
Страница 72: ......
Страница 84: ...System Description 82 Figure 3 7 GemCone nebulizer and end cap N0680343 also shown in cross sec tional view ...
Страница 97: ...Installation 4 ...
Страница 98: ......
Страница 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Страница 164: ...Installation 162 ...
Страница 165: ...Maintenance 5 ...
Страница 166: ......
Страница 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Страница 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 272: ...Maintenance 270 ...
Страница 273: ...Troubleshooting 6 ...
Страница 274: ......
Страница 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Страница 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Страница 296: ...Troubleshooting 294 Figure 6 4 Aux Gas too high Air Leak or Spray Chamber Temperature too high ...
Страница 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Страница 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Страница 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Страница 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Страница 308: ......